Use TPP offset
Menu
>
Select or deselect the use of pressure offset to stabilize the eardrum. To change the
current setting, select
Menu
>
More settings..
>
TPP offset
.
Off
: TPP is used if available
On
: TPP + TPP offset (according to TPP sign) are used
Default value:
On
TPP offset
Menu
>
More settings..
Select a pressure offset to stabilize the eardrum. This may facilitate reflex recordings
from ears showing steeply sloped tympanograms. A positive value offsets in same dir-
ection as TPP sign.
-50 < x < 50 daPa (offset according to TPP sign)
Default value: 20 daPa
Pump during stim(ulation)
Menu
>
More settings..
On
/
Off
If
On
, the pump will actively keep the TPP pressure even during stimulus present-
ation.
If
Off
, the pump will adjust the pressure between presentations.
Default value:
Off
(pump kept inactive during the stimulus period)
Reflex criterion
Menu
>
More settings..
Enter the appropriate setting for the reflex criterion.
The dashed horizontal line in each reflex graph indicates the predefined reflex cri-
terion.It is offset from the base line marker on the vertical axis. The degree of offset is
influenced by this setting.
The applicable unit (mmho, cc, ml, cm
3
or
µ
l) follows unit and probe tone settings.
Default criterion: ±0.02 mmho
Deflection scale
Menu
>
More settings..
Defines the size of the admittance scale for reflex measuring.
The applicable unit (mmho, cc, ml, cm
3
or
µ
l) follows unit and probe tone settings.
If
µ
l is used for 226 Hz probe tone, then all values are multiplied by 1000.
Default: -0.15 mmho for 226 Hz probe tone
Default: -0.30 mmho for 1000 Hz probe tone
Show 500 Hz
Menu
>
More settings..
Select the stimulus type to be included in manual threshold testing.
Default:
On
Show 1000 Hz
Menu
>
More settings..
Select the stimulus type to be included in manual threshold testing.
On
/
Off
Default:
On