A.
Selected test setting
B.
Selected test and frequency
The initial reflex screening test screen with ipsilateral 1
kHz stimulus highlighted.
As indicated by the
Start
softkey, the screening is auto-
matically performed for all preset stimuli.
Reflex Screening test result, where the result field for con-
tralateral BBN is highlighted, and the curve and values for
this stimulus type are shown. Reflexes are found for 1 kHz
contralateral stimulus and BBN ipsilateral stimulus. Reflexes
are absent for 1 kHz ipsilateral and BBN contralateral stim-
uli.
Reflex screening results
The deflection curves of the tested intensities are displayed in the graphs for the highlighted stimulus. Individual results
for stimulus intensity and admittance change are listed below each graph.
•
Check-marks or dashes indicate whether a reflex has been detected or not.
This applies when the measurement setting
Menu
>
More settings..
>
No. of intensities
is set
to
1
.
/
•
Numerical values or crossed out numerical values indicate whether a reflex has been detected
or not.
This applies when the measurement setting
Menu
>
More settings..
>
No. of intensities
is set
to
more than 1
.
•
Blank fields indicate that the test has not been performed.
Changing view settings
No view settings apply to the Reflex Screening test.