3. If contralateral stimulation is used in the test, make sure that both the ipsilateral probe and the contralateral phone
are in place before you start the measurement.
4. If needed, toggle the
Ear Selector
to select the ear on which you wish to start the test.
During testing
Warning
•
If the patient is troubled by the test, stop the test. The test is interrupted and
the pump pressure is relieved immediately.Already measured results are kept.
or
11.7.1
Reflex Decay - procedure
Procedure
1. Press
PageDown
to go to the Reflex Decay test
screen, or select the test from
View
>
Immittance
.
Warning
•
If the patient is troubled by the test, stop the test. The test is interrupted and
the pump pressure is relieved immediately.Already measured results are kept.
or
Automatic RD testing
1. Click
Start
in the
Auto
section of the Control Panel to start a fully automatic reflex decay meas-
urement for the preset stimuli.