Show 1000 Hz
Menu
>
More settings..
Select the stimulus type to be included in reflex decay testing.
Default:
On
Pre-stim(ulus)
time
Menu
>
More settings..
Defines the recording time before each stimulus is presented.
Default value: 1.0 seconds
Stim(ulus) time
Menu
>
More settings..
Defines the duration of the stimulus.
Default value: 5.0 seconds
Post-stim(ulus)
time
Menu
>
More settings..
Defines the recording time after each stimulus is terminated.
Default value: 1.0 seconds
Stimulus1
Menu
>
More settings..
>
Auto Settings..
500 Hz, 1000 Hz.
Default: 500 Hz
Stimulus2
Menu
>
More settings..
>
Auto Settings..
500 Hz, 1000 Hz,
Off
.
Default: 1000 Hz
Stimulus side
Menu
>
More settings..
>
Auto Settings..
I, C, I - > C, C - > I
(Ipsi, Contra, Ipsi then Contra, Contra then Ipsi)
Default: I(psi)
Pause-time
Menu
>
More settings..
>
Auto Settings..
Defines the inter-stimuli interval.
Default value: 0.5 seconds
Note
•
The
Intensity increment
setting used in Reflex Decay is set in the Acoustic Reflex
Threshold setup. This is because the Reflex Decay stimulus level defaults to the reflex
thr10 dB.
Warning
•
When testing manually, be careful when you select the setting for Stimulus
Intensity so as not to expose the patient to too high intensity levels!
•
Whenever a stimulus level exceeds the warning level (> 108 dB SPL re 2cc), the stimulus
intensity value will start flashing and all automatic testing is paused. You are then prompted
to decide whether to continue or to move on to the next stimulus type.
•
In Manual testing, whenever a stimulus level exceeds the Intensity Safety level
((> 115 dB SPL re 2cc), the stimulus intensity value will start flashing.