The dashed horizontal line in each reflex graph indicates the predefined reflex criterion.It is offset from the base line
marker on the vertical axis.
Changing view settings
You can change the following view settings to customize your view. None of these settings influence the test data as such.
If there is test data for a frequency, it will be shown anyway.
•
Menu
>
More settings..
>
Deflection scale
•
Menu
>
s
>
–
Show 500 Hz-
–
Show 1000 Hz-
–
Show 2000 Hz-
–
Show 3000 Hz-
–
Show 4000 Hz-
–
Show BBN-
–
Show LBN-
–
Show HBN
8.8.4
Settings - Reflex Threshold
Menu selection
Setting
Menu
>
More settings..
>
Reflex criterion
Influences the degree of offset.
Automatic and semi-automatic testing
The following settings are useful:
Menu selection
Setting
Menu
>
Procedure options..
>
Auto resume on seal
•
Off
If there is a probe leak, the test is interrupted. You must resume testing manually
if testing is paused due to a leak.
On
Testing resumes automatically when a proper seal is obtained if testing was paused
due to a leak.
Menu
>
Custom TPP offset
•
On
The pressure is offset from the TPP by the TPP offset defined in
Menu
>
More set-
tings..
>
TPP offset
. If the offset is set to a value with the post fix "more", it shifts
the pressure in the same direction as the sign of the TPP, and in the opposite dir-
ection if it is set to "less".
Note
•
The TPP value from the most recent tympanometric curve at the
same probe tone is used in reflex measurements.