13.3.2
ETF-I setup
The following settings are useful:
Layered curves
Menu
>
Set to
On
to enable a typical ETF-I setting with layered tympanograms.
Note
•
This setting is not saved as part of a measurement setup.
Auto next curve
(when
starting new measurement)
Menu
>
More settings..
Select
On
or
Off
to define whether the system will automatically select the next curve
number available for the new measurement results.
Sweep direction
Menu
>
More settings..
Make sure that the same sweep direction is used for all the curves to be compared.
13.4
Acoustic Reflex Screening setup
The following settings are useful:
Auto start on seal
Menu
>
Procedure options..
>
Auto start on seal
Typically for use with screening eartips.
Recommended setting:
Sequence
, with T+RS sequence selected.
Auto resume on seal
Menu
>
Procedure options..
>
Auto resume on seal
Off
: The screening eartip must remain in place against the patient’s ear. If the probe is
removed, the test is interrupted.
On
: If the screening eartip is removed from the patient’s ear, the test continues from the
intensity it came to or was interrupted in, when there is seal.
Probe Tone
Menu
Scroll to select the appropriate frequency for this measurement setup.
Adult default: 226 Hz
Infant default: 1000 Hz
If 226 Hz is selected, you can set admittance to be shown in the admittance unit mmho
or in the compliance volume units ml, cc, cm
3
or
µ
l.
For frequencies higher than 226 Hz, admittance is shown
only
in mmho. The value in
Admittance scale
automatically changes to mmho.
•
226 Hz
•
1000 Hz
Recommended for testing on infants younger than 4-6 months. See also
.
Stimulus1 - 5
Menu
>
More settings..
Select the stimulus types to be included. See the following.