Ipsilateral Stimulation
Tone:
500 Hz, 1000 Hz, 2000 Hz, 3000 Hz, 4000 Hz
Frequency accuracy:
± 0.5%
Noise
White Noise according to IEC 1027
Low Pass 400 to 1600 Hz
High Pass 1600 to 4000 Hz
Roll off >12 dB/Octave
Step size dB:
1, 2, 5, 10 dB
Range at:
500 Hz at 50 to 105 dB HL ± 3 dB
1000 Hz at 50 to 120 dB HL ± 3 dB
2000 Hz at 50 to 115 dB HL ± 3 dB
3000 Hz at 50 to 105 dB HL ± 3 dB
4000 Hz at 50 to 110 dB HL ± 3 dB
THD:
< 3% in 2 cc (measured 5 dB below max output)
Air pressure system
Range:
200 to -400 daPa/s,
Ex400 to -600 daPa/s
Pressure sweep rate:
50, 100, 200, 400 daPa/s, A.F.A.P
A.F.A.P. will start at 500 daPa/s and slow down to 400 daPa, when at peak is detec-
ted.
Pressure accuracy:
±10% or ±10 daPa, whichever is greatest
Pump measure direction:
Positive to negative or negative to positive
Safety:
Separate 530 daPa and -730 daPa. ±70 daPa
Software 450 daPa and -650 daPa. ±70 daPa.
Unit of admittance graph Y-axis
ml, cc, mmho, µl
Unit of graph X-axis
daPa, s
Display
Graphic 128x128 dots
BT antenna
BT antenna:
Chip multilayer antenna for 2.4 GHz
Antenna gain:
2 dBi
Antenna impedance:
50 Ohm
Otometrics - MADSEN OTOflex 100
177
19 Technical specifications