Manual RD testing
1. Check and, if needed, adjust the Control Panel and Test settings.
–
Select the settings needed for
Stimulus
,
Intensity
and
Stimulus side
in the
Manual
sec-
tion of the Control Panel.
If you select a
Control Panel
>
Stimulus
setting not included in your view setting setup,
the Thresholds table is automatically updated to show the selection when the test is star-
ted.
–
Custom TPP offset
in the
Reflex Threshold
section of the Control Panel:
Checked
: If this setting is applied in the Reflex Threshold test, it is automatically included
in the Acoustic Reflex Decay test, but not vice versa.
–
Deflection axis
-
Scale
(
Tools
>
Options
>
View
):
If needed, adjust this view setting.
2. In the Reflex Decay screen, click
Start
on the Control Panel.
3. If needed, repeat testing of more stimulus levels.
Pausing the test
1. To interrupt the test, click/press
Pause
.
2. To resume the test, click/press
Pause
again. After the pause, the test resumes at the intensity
it came to or was interrupted in.
Continuing the test
1. When you have finished testing one ear, switch ears, if needed.
2. To continue testing, click/press
Start
and test the other ear.
The test progresses as follows
Deflection graph
The deflection graph is limited by blue markers indicating stimulus on and off.
A third marker appears as indication of the half-life time where the curve decreases to 50% of its
initial deflection.
110
Otometrics - MADSEN OTOflex 100
11 Testing with the OTOsuite Immittance module