Pressure control
in ETF-P.
5.3
The Tympanometry screen
The Tympanometry screen
A. Tymp results, showing:
–
Probe tone
–
TPP (Tympanometric Peak Pressure)
–
SA/SC (Static Admittance/Static Compliance)
–
TW (Tympanometric Width)
–
ECV (Ear Canal Volume)
–
Type, if selected (if normal region and baseline are
enabled).
B. Norm area, if selected
C. The tympanometric curves
Scaling the tympanogram
Auto Scale
Click to select/deselect autoscaling of a tympanogram in order to display the entire curve. When
you change the ear or the patient, the scale will revert to the default scale.
When no data is available, the scale selected will be used.
Note
•
Autoscaling adjusts viewing dynamically according to the highest of all currently displayed curves, so that
you can visually compare ear results.
Viewing the tympanogram
The tympanogram can be viewed in four different ways
•
traditional total admittance, Y
•
traditional baseline compensated admittance, Y
tm
•
susceptance, B, and conductance, G; both total and compensated
•
component compensated admittance, Y