Procedure
During the measurement, two diamond markers are shown:
–
an admittance marker just to the right of the admittance axis, indicating the current admit-
tance
–
a pressure marker below the pressure axis, indicating the current pressure.
You can save up to 3 separate measurements for each ear.
Pausing the test
–
To interrupt the test, click/press
Pause
.
–
To resume the test, click/press
Pause
again. After the pause, the test resumes at the
intensity it came to or was interrupted in.
Stopping the test
–
Press to stop test.
2. To make a new sweep, Click
Start
.
You can save up to 3 separate measurements for each ear.
3. When you have finished testing one ear, switch ears, if needed.
/
4. To continue testing, press
Start
and test the other ear.
5. To view tympanometry results, see
The diagnostic tympanometry result
.
6. When you have finished testing, see
Recommended settings - Diagnostic tympanometry
Setting
Value
Toolbar icon
Auto next curve...
(
Tools
>
Options
>
Tymp. and Reflex Scr.
> )
The test starts automatically as soon as a good probe fit with an airtight
seal is achieved. Regardless of the setting in
Tools
>
Options
>
Tymp.
and Reflex Scr.
>
Auto next curve...
, the test automatically changes to
the next tymp curve for each tymp sweep. This is to prevent overwriting
already recorded tymps, if a tymp measurement is triggered by accident.
On
Pressure axis - Range
(
Tools
>
Options
>
Tymp. and Reflex Scr.
> )
The pump increases the pressure to the maximum value set here. This
value depends on the value set in
Sweep direction
.
Sweep direction
(
Control Panel
)
The air pressure determines the direction of the pressure sweep from a
positive value to a negative value or vice versa.
Default:
Negative
.
98
Otometrics - MADSEN OTOflex 100
11 Testing with the OTOsuite Immittance module