14.3.5.2
Reflex Decay: Test settings
Click icon or select
Tools
>
Options
>
Reflex Decay
>
Measurement
.
General
Pump during stim-
ulus
On
/
Off
Default value:
Off
Custom TPP offset
(daPa)
Select a pressure offset to stabilize the eardrum. This may facilitate
reflex recordings from ears showing steeply sloped tympanograms.
The TPP offset can be adjusted in increments of
xx daPa more/less
.
50 daPa more to 50 daPa less (offset according to TPP sign)
Default value: 35 daPa
Stimulus side
Adult default: Ipsi
Infant default: Ipsi
•
Ipsi
Stimulus is presented in the probe ear where the measurements
are made.
•
Contra
Stimulus is presented through an ear phone or insert phone in
the ear opposite to the probe ear.
•
Ipsi then Contra
•
Contra then Ipsi
Stimulus Timing
(sec.)
Pre-stim(ulus time)
Defines the recording time before each stimulus is presented.
Default value: 0.2 seconds
Stimulus (time)
Defines the duration of the stimulus. Displays the text
Manual
when
set to 0.
Default value: 1.3 seconds
Post-stim(ulus time)
Defines the recording time after each stimulus is terminated.
Default value: 0.5 seconds
Pause (time)
Defines the inter-stimulus interval.
Default value: 0.5 seconds
Stimulus types
(Auto-
matic)
Stimulus types
Select the stimulus types to be included/excluded. At least one stim-
ulus type must always be selected.
Available types: 500 Hz, 1000 Hz.
Otometrics - MADSEN OTOflex 100
159