Setting
Value
Auto resume on seal
(
Menu
>
Procedure options..
>)
This function is ideal for speeding up the test procedure.
–
On
:
If the screening eartip is removed from the patient’s ear, the test continues from the
intensity it came to or was interrupted in, when there is seal.
–
Off
:
The screening eartip must remain in place against the patient’s ear. If the probe is
removed, the test is interrupted.
On
or
Off
Stop when results are available
(
Menu
>
More settings..
>)
The measurement stops automatically when satisfactory values have been registered.
On
Max. intensity
(
Menu
>
More settings..
>
To avoid automatic testing being interrupted because of high stimulus intensity levels when reach-
ing the warning limits, it is recommended that you set the max. intensity to 95 dB HL.
95 dB HL
No. of intensities
(
Menu
>
More settings..
>
–
1 intensity
.
Check marks
or dashes
indicate whether a reflex has been detected or not.
–
2 or 3 intensities
.
Numerical values or crossed out numerical values indicate whether a reflex has been
detected or not.
Intensity increment
(
Menu
>
More settings..
>
Step in dB between each stimulus intensity presented. If e.g. the increment is set to 10 dB and the
number of intensities is set to 2, the first presentation will be at the max intensity -10 dB.
8.4.3
The reflex screening result
1. Press
On
/
Off
briefly to select the Reflex Screening test screen.
The deflection curves of the included intensities are shown in the graphs for the highlighted stimulus. Individual stim-
ulus intensity and admittance change are listed below each graph.
Otometrics - MADSEN OTOflex 100
65
8 Testing with MADSEN OTOflex 100