Stimulus side
Adult default: Ipsi
Infant default: Ipsi
•
Ipsi
Stimulus is presented in the probe ear where the measurements
are made.
•
Contra
Stimulus is presented through an earphone or insert phone in
the ear opposite to the probe ear.
Auto
Stimulus side
•
Ipsi
Stimulus is presented in the probe ear where the measurements
are made.
•
Contra
Stimulus is presented through an ear phone or insert phone in
the ear opposite to the probe ear.
•
Ipsi then Contra
•
Contra then Ipsi
14.3.4.2
Reflex Threshold: Test settings
Click icon or select
Tools
>
Options
>
Reflex Threshold
>
Measurement
General
Pump during stim-
ulus
On
/
Off
Default value:
Off
Custom TPP offset
(daPa)
Select a pressure offset to stabilize the eardrum. This may facilitate
reflex recordings from ears showing steeply sloped tympanograms.
The TPP offset can be adjusted in increments of
xx daPa more/less
.
50 daPa more to 50 daPa less (offset according to TPP sign)
Default value: 35 daPa
Reflex criterion
(mmho)
Enter the appropriate setting for the reflex criterion. It follows unit
and probe tone settings.
Default: ±0.02 mmho
Stimulus Intensities
You can select the start and maximum stimulus intensities in fields directly below
Stimulus
Intensities
.
Whenever the stimulus intensity level exceeds 108 dB SPL, for example when you select
100 dB HL at 500 Hz Ipsi, a warning will appear, and the automatic test is paused. You will be
prompted whether to continue or skip to the next stimulus type.
To avoid automatic testing being interrupted because of high stimulus intensity levels when
reaching the warning limits, it is recommended that you set the max. intensity to 95 dB HL.
156