4. If contralateral stimulation is used in the test, make sure that both the ipsilateral probe and the contralateral phone
are in place before you start the measurement.
5. Ask the patient to sit very still and quiet during the test, without moving head or jaw.
6. If needed, toggle the
Ear Selector
to select the ear on which you wish to start the test.
7. Load a test setting configured for screening patients in the relevant age group.
8. Continue with either
or
.
11.2.1
Screening tympanometry
The default test setting applying to the
Tympanometry
test screen is not designed specifically for screening tym-
panometry.
Procedure
The test starts automatically with tympanometry testing and continues automatically to reflex screening. (if Reflex
Screening is selected)
1. In order to prepare MADSEN OTOflex 100 to start measuring as soon as seal is achieved, do the following:
2. Load a test setting configured for screening patients in the relevant age group. See recommended settings below.
3. If needed, toggle the
Ear Selector
to select the ear on which you wish to start the test.
Screening tympanometry
–
Without applying the eartip to the patient’s ear click
Start
in the
Tympanometry
section
of the Control Panel.
Sequence testing
–
In the
Tympanometry
screen, select
T+RS
in the
Sequence
field below the Control
Panel.
–
Without applying the eartip to the patient’s ear click
Start
in the
Sequence
section of the
Control Panel.
4. Apply the probe to the patient’s ear with a steady grip. The test starts automatically as soon as
a good probe fit with an airtight seal is achieved.
Warning
•
If the patient is troubled by the test, stop the test. The test is interrupted and
the pump pressure is relieved immediately.Already measured results are kept.