Setting
Value
Toolbar icon
Increment (dB)
(
Tools
>
Options
>
Tymp. and Reflex Scr.
>
Stimulus
Intensities
>)
•
Step in dB between each stimulus intensity presented. If e.g. the
increment is set to 10 dB and the number of intensities is set to 2,
the first presentation will be at the max intensity
-10 dB.
To avoid automatic testing being interrupted because of high stim-
ulus intensity levels when reaching the warning limits, it is recom-
mended that you set the max. intensity to 95 dB HL. .
Reflex Criterion
(
Tools
>
Options
>
Tymp. and Reflex Scr.
>
Auto
Reflex Screening
> )
Reflex Screening results
The Reflex results shown in the Reflex Threshold view are reflex screening results. If, subsequently a Reflex Threshold test
is made using the same stimulus types as in Reflex Screening, the results from this test will replace the screening values
shown in this view.
The deflection curves of the tested intensities are displayed in the graphs for the highlighted stimulus. Individual results
for stimulus intensity and admittance change are listed below each graph.
•
Check-marks or dashes indicate whether a reflex has been detected or not.
This applies when the measurement setting
No. of intensities
is set to
1
.
/
•
Numerical values or crossed out numerical values indicate whether a reflex has been
detected or not.
This applies when the measurement setting
No. of intensities
is set to
more than 1
.
•
Blank fields indicate that the test has not been performed.
11.3
Tympanometry
Note
•
It is recommended that you perform a tympanometric test before making any acoustic reflex meas-
urement, and determine the acoustic reflex threshold before making a reflex decay measurement.
96
Otometrics - MADSEN OTOflex 100
11 Testing with the OTOsuite Immittance module