Manual RT testing
1. Check and, if needed, adjust the Control Panel and Test settings.
–
Select the settings needed for
Stimulus
,
Intensity
and
Stimulus side
in the
Manual
section of the Control Panel.
If you select a
Control Panel
>
Stimulus
setting not included in your view setting
setup, the Thresholds table is automatically updated to show the selection when the
test is started.
–
Custom TPP offset
in the
Reflex Threshold
section of the Control Panel:
Checked
: This setting is applied automatically in the Acoustic Reflex Decay test set-
ting, but not vice versa.
–
Deflection axis
-
Scale
(
Tools
>
Options
>
View
):
If needed, adjust this view setting.
2. In the Reflex Threshold screen, click
Start Stimulus
in the
Manual
section of the Control
Panel to present a single stimulus intensity.
3. To manually determine a threshold, click on the drop-down box of the specific field in the
Thresholds table and select the stimulus intensity to be reported as the threshold value.
Note
•
The audiogram symbol for
Not found
will remain until a threshold is determ-
ined manually.
If a measurement consists of more than 3 intensities, toggle the graph selector to view the
graphs.
4. If needed, repeat testing of more stimulus levels.
Pausing the test
1. To interrupt the test, click
Pause
on the Control Panel.
2. To resume the test, click
Pause
again. After the pause, the test resumes at the intensity it
came to or was interrupted in.
Continuing the test
1. When you have finished testing one ear, switch ears, if needed.
2. To continue testing, click
Start
and test the other ear.
104
Otometrics - MADSEN OTOflex 100
11 Testing with the OTOsuite Immittance module