14.3.2.2
ETF-I: Test settings
Click icon or select
Tools
>
Options
>
Tymp. and Reflex Scr.
>
Measurement
No special settings are needed for ETF-I tests, except for the following:
Auto next curve...
Enable
Auto next curve...
to facilitate recording of multiple tym-
panograms.
14.3.3
Reflex Screening setup
When you perform a Reflex Screening measurement, there are a number of settings you can adjust. These settings are
available either in the Control Panel on the left side of the screen, or in the
Tools
>
Options
>
Tymp. and Reflex Scr.
>
Measurement
dialog.
14.3.3.1
Reflex Screening: Control Panel settings
Reflex Screening
Probe Tone
Select the frequency for this measurement setup.
226 Hz
Adult default
If 226 Hz is selected, you can set admittance to be shown in the
admittance unit mmho or in the compliance volume units ml, cc or
cm
3
as described in
Tymp. and Reflex Scr. view settings
.
For frequencies higher than 226 Hz, admittance is shown
only
in
mmho. The value in
Y scale
automatically changes to mmho.
1000 Hz
Infant default
Recommended for testing on infants younger than 4-6 months. See
also the MADSEN OTOflex 100 Guide.
14.3.3.2
Reflex Screening: Test settings
Click icon or select
Tools
>
Options
>
Tymp. and Reflex Scr.
>
Measurement
Auto Reflex Screen-
ing
Reflex criterion
(mmho)
Enter the appropriate setting for the reflex criterion. It follows unit
and probe tone settings.
Default: ±0.02 mmho
Otometrics - MADSEN OTOflex 100
153
14 Immittance module settings