7.3.4
Daily probe check
To make sure that the probe is functioning correctly, it is recommended that you perform a probe check at the start of
each day.
Note
•
If your test environment changes, for instance if there is an increase in humidity, or if you are going to test
at an altitude different from the one set in MADSEN OTOflex 100, adjust the altitude setting and make another
probe check.
1. Make sure that the probe tip has been cleaned
and disinfected before you place it in the test
cavity. This is to make sure that the probe tip
and filter do not influence the probe test, and
that the test cavity is not contaminated.
See
Cleaning and disinfecting the probe tip
2. Select
Menu
>
Probe Check
.
3. Insert the probe tip without eartip in the test
cavity in the charger. The probe check starts
automatically. The probe is checked for occlu-
sion and leakage. If the probe is OK, the probe
is automatically calibrated to 2cc.
If there is a probe error
In case of a probe error, check the following:
•
Make sure that the threaded ring holding the probe tip in place is firmly tightened.
•
Make sure that the sound channels in the probe tip are clear and that the probe is connected.
•
Make sure that the probe tip goes into the cavity at a 90° angle.
If the probe is faulty, contact an authorized service department for repair.
7.3.5
Users
If several users are using MADSEN OTOflex 100, you can select a user for the session, or create a new user.
OTOsuite reports show the selected user. The user selected in MADSEN OTOflex 100 is
not
reflected in personalized set-
tings and protocols.
Selecting a user
•
With
OTOsuite:
If you are using MADSEN OTOflex 100 with OTOsuite, you cannot access the
Patient & User
screen in MADSEN
OTOflex 100. Patient and user information is set up in OTOsuite.
•
As a stand-alone device:
If you are using MADSEN OTOflex 100 as a stand-alone device, the list shows only the users you have created in
MADSEN OTOflex 100.
1. Select
Menu
>
Patient & User
, scroll to the
User
field and press
Select
to access the field.
Otometrics - MADSEN OTOflex 100
51