1. To replace the wax filter,
remove the probe tip.
2. Use the extraction pin to
remove the wax filter from the
probe tip.
Do not put used filters in the
accessory box.
Filters are disposable.
3. Insert a new filter. Be careful
not to damage the filter open-
ings.
4. Fit the probe tip over the new
filter in the probe body and
screw the threaded ring back
onto the probe body. Tighten
well.
A.
Wax filter
B.
Extraction lever
15.5
Cleaning MADSEN OTOflex 100
Prerequisites
•
Before cleaning, switch off MADSEN OTOflex 100 and disconnect it from any external power source.
•
Unplug the probe from MADSEN OTOflex 100.
The device
•
Remove dust using a soft brush.
•
Use a soft, slightly damp cloth with a small amount of mild detergent or approved non-caustic medical grade dis-
infectant wipes to clean the unit according to local infection control regulations.
Keep the unit away from liquids. Do not allow moisture inside the unit. Moisture inside the unit can damage the instru-
ment and it may result in a risk of electrical shock to the user or patient.
Clean the device screen periodically. Use an anti-static non-solvent solution on a lint-free cloth.
The probe tip
See
Cleaning and disinfecting the probe tip
.
15.5.1
Eartips
Eartips are in direct contact with your patients, and you should therefore observe strict hygienic precautions to prevent
passing infections from one patient to another.
•
Use only disposable eartips.
•
Dispose of eartips after use.
Disposal
MADSEN OTOflex 100 eartips and foam eartips for the E-A-RTONE® 3A insert phone are disposable, and should not be
cleaned or re-used. Always throw away disposable eartips after use on a patient. See
for
166
Otometrics - MADSEN OTOflex 100
15 Service and maintenance