Data management in MADSEN OTOflex 100
12.1
Data handling
Handling of data shared by MADSEN OTOflex 100 and OTOsuite is to a large degree automated.
As a general rule:
•
Communication between OTOsuite and MADSEN OTOflex 100 is defined by MADSEN OTOflex 100. This means that
the patient folder currently shown on MADSEN OTOflex 100 will be transferred automatically to OTOsuite.
•
Measurements settings are always synchronized, allowing MADSEN OTOflex 100 to continue unaffected if the con-
nection to the PC should be lost.
12.1.1
Interruption of communication
Communication via Bluetooth is in MADSEN OTOflex 100 set to
On
as default.
If the Bluetooth connection between MADSEN OTOflex 100 and OTOsuite is interrupted, communication with MADSEN
OTOflex 100 is disconnected.
•
If reconnecting while the current patient folder is the same in MADSEN OTOflex 100 and OTOsuite, synchronization is
done automatically.
•
If reconnecting after the current patient folder is changed on either MADSEN OTOflex 100 or in OTOsuite, you will be
prompted to synchronize data or cancel.
12.1.2
Synchronization
Data is usually transferred from MADSEN OTOflex 100 to OTOsuite and the data is never deleted without confirmation
from the user.
Synchronization takes just a few seconds and is in many cases automatically initiated on connection.
•
If you synchronize, the patient folder shown in MADSEN OTOflex 100 will be overwritten.
•
If you cancel synchronization, you can save/print out the patient folder shown in MADSEN OTOflex 100, and then syn-
chronize.
Note
•
Only the patient folder currently shown in the test device will be transferred to OTOsuite during syn-
chronization. Other patient folders made while the test device was off-line must be transferred manually. To do
so, see
Uploading test results to OTOsuite
Synchronization of corrections made to measurements
If you have made changes to for instance reflex threshold or tympanometric classification values, either in OTOsuite or in
MADSEN OTOflex 100, these changes are given a time stamp.
The most recent change, regardless of whether it was made in OTOsuite or in MADSEN OTOflex 100, will be applied or
synchronized.
Synchronizing patient folders
If patient folders on the PC and in the test device are not synchronized when OTOsuite connects to MADSEN OTOflex 100,
they will be synchronized automatically or you will be prompted to resolve synchronization conflicts, depending on the set-
tings in the Configuration Wizard.
118
Otometrics - MADSEN OTOflex 100
12 Data management in MADSEN OTOflex 100