4. Click
Pause
to manually pause the progress of the sequence.
When you resume reflex testing, the test continues at the intensity level during which the interruption
occurred.
Click again to resume testing.
5. Click
Stop
to stop the sequence manually.
Skipping a measurement
6. Click
Skip
to skip the ongoing measurement and continue with the next.
Resuming a sequence
•
If
Tools > Options > General > Auto resume on seal
is
On
, and a leak is detected or the probe is removed from the
patient's ear, the sequence pauses automatically. The sequence will automatically resume from the pause state when
seal has been reestablished.
The sequence continues from where it was paused. An interrupted reflex stimulus will be repeated.
Note
•
If the leak occurs during the tympanometric measurement, the tympanometry part of the sequence will
not be repeated. Press
Stop
and restart the sequence if you need to repeat the tympanometric measurement.
•
If
Tools > Options > General > Auto resume on seal
is
Off
:
The probe must remain in place throughout the sequence. If the probe is removed, the test will restart with Tym-
panometry testing when testing is resumed.
11.2
Screening
You can perform Tympan Reflex Screening (T+RS) as a sequence from the Tympanometry test screen.
Note
•
The safety intensity level cannot be exceeded in screening mode.
Note
•
You can use either ordinary eartips or screening eartips for screening.
If you are using screening eartips and the setting
Auto start on seal
is set to
On
, the test will start with tym-
panometry testing and continue automatically from tympanometry to reflex screening when seal is obtained, i.e.
when the screening eartip is pressed gently against the ear.
Preparing for testing
1. Do as described in
.
2. Make sure that MADSEN OTOflex 100 is switched on.
3. Launch the Immittance module with the
Tympanometry
test selected.
Otometrics - MADSEN OTOflex 100
91
11 Testing with the OTOsuite Immittance module