3. If you are cleaning the probe
tip
during a session testing a
patient
, use the cleaning brush
to clean the cleaning wire,
especially where it protrudes
from the probe tip.
If you are cleaning the probe
tip
between sessions
, use dis-
infectant to clean the cleaning
wire, and, if needed, the brush.
Caution
•
Even the slightest
amount of moisture may dis-
solve any residual cerumen
and thus contaminate the
sensitive parts in the body of
the probe.
A.
Cleaning wire
Caution
•
Never clean the
sound channels in the probe
body, as this may cause dam-
age to the probe.
4. Make sure that the sound chan-
nels are completely dry before
fitting the tip back into the
probe body, or use a spare
probe tip.
5. Fit the probe tip and screw the
threaded ring firmly back onto
the probe body. Tighten well
to prevent leakage.
15.4.1
Cleaning and disinfecting the probe tip
Ear canal debris blocking the probe tubes can lead to abnormally large ear canal volume readings, leak messages, and other
odd results. Check the channels of the probe tip every time you use the probe. Even small amounts of cerumen or vernix
can block the probe channels.
Caution
•
Thorough cleaning of the probe tip is required after use in infected ear canals. Cleaning the threaded
ring may also be necessary.
Methods
The probe tip material is highly resistant to a wide range of temperature and chemical influences.
164
Otometrics - MADSEN OTOflex 100
15 Service and maintenance