tympanometry testing and continue automatically from tympanometry to reflex screening when seal is obtained,
i.e. when the screening eartip is pressed gently against the ear.
Preparing for testing
1. Do as described in
.
2. If contralateral stimulation is used in the test, make sure that both the ipsilateral probe and the contralateral phone
are in place before you start the measurement.
3. Ask the patient to sit very still and quiet during the test, without moving head or jaw.
4. If needed, toggle the
Ear Selector
to select the ear on which you wish to start the test.
8.4.1
Screening tympanometry
A.
Press to start the test
Procedure
The test starts automatically with tympanometry testing and continues automatically to reflex screening. (if Reflex
Screening is selected)
Prepare MADSEN OTOflex 100 to start measuring as soon as seal is achieved:
1. Do as described in
.
2. Load a test setting configured for screening patients in the relevant age group. See recommended settings below.
3. If needed, toggle the
Ear Selector
to select the ear on which you wish to start the test.
Screening tympanometry
–
Use the Test Selector to select the
Tympanometry
screen.
–
Without applying the eartip to the patient’s ear press the right-hand softkey.
Sequence testing
The test starts automatically with tympanometry testing and continues automatically to reflex
screening.
–
Sequence, select
Menu
>
Procedure options..
>
T+RS
.
–
Without applying the eartip to the patient’s ear press the right-hand softkey.
62
Otometrics - MADSEN OTOflex 100
8 Testing with MADSEN OTOflex 100