![otometrics MADSEN OTOflex 100 Скачать руководство пользователя страница 75](http://html.mh-extra.com/html/otometrics/madsen-otoflex-100/madsen-otoflex-100_reference-manual_1658679075.webp)
Semi-automatic RT testing
–
Use the
Scroll wheel
to select the stimulus (frequency and ipsi/contra) to be tested.
–
Press the right-hand softkey to start a semi-automatic reflex threshold search for the selec-
ted stimulus.
The test runs a search for the selected stimulus.
Manual RT testing
If a threshold is not found during automatic or semi-automatic testing, you can test specific stimulus intensities
manually.
–
Press
Select
to activate manual test mode.
–
Scroll to select the stimulus intensity for a single presentation.
–
Press the right-hand softkey.
This performs a threshold search for the currently highlighted stimulus level.
If needed, repeat testing of more stimulus levels.
–
Mark the intensity with which you associate the threhold.
–
Adjust the intensitiy and press the right-hand softkey to present the stimulus.
–
Continue until you register the reflex threshold or until the max. intensity is reached.
The measurement stops automatically at the end of each stimulus presentation.
–
To test another level, use the scroll wheel to scroll to the next stimulus level and press
the right-hand softkey to present the stimulus.
–
Use the scroll wheel to scroll to the stimulus level to be reported as the threshold value,
and mark the intensity.
Pausing the test
•
To interrupt the test, click/press
Pause
.
•
To resume the test, click/press
Pause
again. After the pause, the test resumes at the intensity
it came to or was interrupted in.
Continuing the test
2. When you have finished testing one ear, switch ears, if needed.
3. To continue testing, click/press
Start
and test the other ear.
Otometrics - MADSEN OTOflex 100
75
8 Testing with MADSEN OTOflex 100