5-28 Image Optimization
Dual Live
In live mode or freeze mode, tap [Dual Live] to enable the dual live function. Both the contrast
image (marked with “
”) and the tissue image (marked with “
”) are displayed (the position can
be changed).
Tips:
In dual live mode, the screen displays the contrast image and tissue image
In freeze mode, only one cine review progress bar displays as the contrast image and
tissue image are reviewed synchronously.
Image position
Use [CEUSPos: XX] to adjust the position of the contrast image.
When it is selected as “Left”, the image displays at the left part of the image area on the screen.
Mix Map
This function combines the contrast image with the tissue image, so that the contrast regions of
interest can be located.
Select options besides [Mix] to select different mixing modes: Contrast and C&T.
When the dual live function is on, you can see the mixed effect on the contrast image.
When the dual live function is off, you can see the mixed effect on the full-screen image.
iTouch
In contrast status, you can also get a better image effect by using the iTouch function.
Tap
to turn the function on (the icon turns blue).
Long press
to exit the function.
Image Saving
Live capture
In live mode, you can save the images of interest by tapping [Pro Capture] and [Retro Capture].
Cine saving
In freeze mode, tap [Save Clip] at the bottom of the operating panel.
Содержание TE5
Страница 1: ...TE7 TE5 Diagnostic Ultrasound System Operator s Manual Basic Volume ...
Страница 2: ......
Страница 6: ......
Страница 12: ......
Страница 24: ......
Страница 36: ......
Страница 54: ......
Страница 56: ...4 2 Exam Preparation 4 1 1 New Patient Information The Patient Info screen is shown as follows 2 1 3 ...
Страница 110: ......
Страница 115: ...Display Cine Review 6 5 6 Tap Return on the screen or tap Freeze to exit image compare ...
Страница 120: ......
Страница 124: ......
Страница 156: ......
Страница 174: ......
Страница 192: ...12 18 Setup Select Advanced and do as follows Select MAPS and do as follows ...
Страница 202: ...13 2 Probes and Biopsy C5 2s L12 4s L7 3s P4 2s L14 6s C11 3s L14 6Ns V11 3Ws P7 3Ts 7LT4s ...
Страница 203: ...Probes and Biopsy 13 3 7L4s P10 4s L20 5s P7 3s L14 5sp SC6 1s SP5 1s 6CV1s L9 3s C5 1s L11 3VNs C4 1s ...
Страница 222: ...13 22 Probes and Biopsy No Name Description 8 Grooves of the needle guided bracket Matched with the tabs of the probe ...
Страница 226: ...13 26 Probes and Biopsy NGB 034 NGB 035 ...
Страница 250: ......
Страница 272: ......
Страница 276: ...A 4 Wireless LAN Tap Add Manually create a network profile to set ...
Страница 282: ......
Страница 318: ......
Страница 322: ......
Страница 323: ...P N 046 006959 07 1 0 ...