Probes and Biopsy 13-43
13.2.8 Clean and Sterilize the Needle-Guided Bracket
Cleaning
Follow the cleaning instructions in the manual.
1. Wear sterile gloves to prevent infection.
2. Wash the needle-guided bracket with water or soapy water to remove all external material. Or,
clean the needle-guided bracket with a urethane sponge.
3. After washing, dry the needle-guided bracket using sterile cloth or gauze.
Sterilization
1. Wear sterile gloves to prevent infection.
2. Clean the needle-guided bracket before sterilizing it. MINDRAY recommends the following
solution or sterilizing system for sterilizing the needle-guided bracket.
3. Follow local regulations when selecting and using disinfectant.
Glutaraldehyde-based sterilant:
Chemical
name
Trade name
Procedures
Glutaraldehyde
(2.2-2.7%)
Cidex
Activated
Glutaraldehyde Solution
Refer to the instructions provided by the solution
manufacturer for details.
The plastic bracket can be sterilized using Cidex Activated Glutaraldehyde Solution at least 233
times (10 hours per time) before its safety and performance is affected.
Hydrogen Peroxide and Peroxyacetic Acid based sterilant:
Trade Name
Chemical Name
Procedures
Minncare
®
Cold
Sterilant
22% Hydrogen Peroxide
4.5% Peroxyacetic Acid
Refer to the instructions provided by the
solution manufacturer for details.
The plastic bracket can be sterilized using Minncare COLD STERILANT at least 245 times (11
hours per time) before its safety and performance is affected.
Refer to the instructions provided by the chemical manufacturer concerning concentration
of the solution, and disinfection and dilution methods. Note that glutaraldehyde disinfectant
solution requires an activating solution.
Rinse the needle-guided bracket thoroughly with sterile water to remove all chemical
residues from it.
After rinsing, dry the needle-guided bracket with sterile cloth or gauze.
STERRAD 100S low-temperature hydrogen peroxide gas plasma sterilization system
Chemical name
Trade name
Procedures
Hydrogen
peroxide gas
plasma
Hydrogen
peroxide vapor
Refer to the instructions provided by the solution
producer for details.
Refer to the STERRAD 100S sterilizing system's instructions provided by the manufacturer
for operation instructions and cautions.
The STERRAD 100S low-temperature hydrogen peroxide gas plasma sterilization system
is available for metal needle-guided brackets.
Содержание TE5
Страница 1: ...TE7 TE5 Diagnostic Ultrasound System Operator s Manual Basic Volume ...
Страница 2: ......
Страница 6: ......
Страница 12: ......
Страница 24: ......
Страница 36: ......
Страница 54: ......
Страница 56: ...4 2 Exam Preparation 4 1 1 New Patient Information The Patient Info screen is shown as follows 2 1 3 ...
Страница 110: ......
Страница 115: ...Display Cine Review 6 5 6 Tap Return on the screen or tap Freeze to exit image compare ...
Страница 120: ......
Страница 124: ......
Страница 156: ......
Страница 174: ......
Страница 192: ...12 18 Setup Select Advanced and do as follows Select MAPS and do as follows ...
Страница 202: ...13 2 Probes and Biopsy C5 2s L12 4s L7 3s P4 2s L14 6s C11 3s L14 6Ns V11 3Ws P7 3Ts 7LT4s ...
Страница 203: ...Probes and Biopsy 13 3 7L4s P10 4s L20 5s P7 3s L14 5sp SC6 1s SP5 1s 6CV1s L9 3s C5 1s L11 3VNs C4 1s ...
Страница 222: ...13 22 Probes and Biopsy No Name Description 8 Grooves of the needle guided bracket Matched with the tabs of the probe ...
Страница 226: ...13 26 Probes and Biopsy NGB 034 NGB 035 ...
Страница 250: ......
Страница 272: ......
Страница 276: ...A 4 Wireless LAN Tap Add Manually create a network profile to set ...
Страница 282: ......
Страница 318: ......
Страница 322: ......
Страница 323: ...P N 046 006959 07 1 0 ...