5-8 Image Optimization
Line Density
Description
The function determines the quality and information of the image.
Operations
Adjust through [Line Density] on the menu.
There are four levels of line density available: UH, M, H, L.
Effects
The higher the line density, the higher the resolution.
Impacts
The higher the line density, the lower the frame rate.
Dynamic Range
Description
This function is used to adjust the B image resolution to compress or expand the
gray display range. The real-time value displays in the image parameter area in
the top-left corner of the screen.
Operations
Adjust through [Dyn Ra.] on the menu.
Tap [-] or [+] to change the value slightly or drag the control directly.
The adjusting range is 30-240, in increments of 5.
Impacts
The more the dynamic range, the more specific the information and the lower
the contrast with more noise.
iClear
Description
The function is used to enhance the image profile so as to distinguish the image
boundary for optimization.
Operations
Adjust through [iClear] on the menu.
Tap [-] or [+] to change the level slightly or drag the control directly.
The system provides 7 levels of iClear adjustment: off represents no iClear
effect, and the bigger the value the stronger the effect.
Impacts
It may cause increased noise.
Persistence
Description
This function is used to superimpose and average adjacent B images, so as to
optimize the image and remove noise.
Operations
Adjust through [Persistence] on the menu.
Tap [-] or [+] to change the level slightly or drag the control directly.
The system provides 7 levels of frame average adjustment: the bigger the value
the stronger the effect.
Effects
Persistence can remove image noise to make details clearer.
Impacts
Increasing Persistence may lead to missing signals.
TSI (Tissue Specific Imaging)
Description
The TSI function is used to optimize the image by selecting acoustic speed
according to tissue characteristics.
Operations
Select different TSI modes using the [TSI] control.
The system provides four ways of optimizing for specific tissues: general,
muscle, fluid and fat.
Содержание TE5
Страница 1: ...TE7 TE5 Diagnostic Ultrasound System Operator s Manual Basic Volume ...
Страница 2: ......
Страница 6: ......
Страница 12: ......
Страница 24: ......
Страница 36: ......
Страница 54: ......
Страница 56: ...4 2 Exam Preparation 4 1 1 New Patient Information The Patient Info screen is shown as follows 2 1 3 ...
Страница 110: ......
Страница 115: ...Display Cine Review 6 5 6 Tap Return on the screen or tap Freeze to exit image compare ...
Страница 120: ......
Страница 124: ......
Страница 156: ......
Страница 174: ......
Страница 192: ...12 18 Setup Select Advanced and do as follows Select MAPS and do as follows ...
Страница 202: ...13 2 Probes and Biopsy C5 2s L12 4s L7 3s P4 2s L14 6s C11 3s L14 6Ns V11 3Ws P7 3Ts 7LT4s ...
Страница 203: ...Probes and Biopsy 13 3 7L4s P10 4s L20 5s P7 3s L14 5sp SC6 1s SP5 1s 6CV1s L9 3s C5 1s L11 3VNs C4 1s ...
Страница 222: ...13 22 Probes and Biopsy No Name Description 8 Grooves of the needle guided bracket Matched with the tabs of the probe ...
Страница 226: ...13 26 Probes and Biopsy NGB 034 NGB 035 ...
Страница 250: ......
Страница 272: ......
Страница 276: ...A 4 Wireless LAN Tap Add Manually create a network profile to set ...
Страница 282: ......
Страница 318: ......
Страница 322: ......
Страница 323: ...P N 046 006959 07 1 0 ...