Measurement 7-3
Gynecology
measurements - used for the uterus, ovary and follicles, etc.
Small Part measurements - used for small parts such as the thyroid, breast, MSK, superficial.
Urology measurements - used for the prostate, renal, micturated and testicular volume.
Vascular measurements - used for carotid, cerebral, upper and lower extremities vessels, etc.
EM measurements - used for EM-related full functional measurements.
Nerve measurements - used for nerve measurements.
7.4
Measurement Accuracy
Table 1 Basic Dimension Measurements
Parameter
Range
Error
Distance
Full Screen
Within ±4%
Area (ellipse, circle) Full Screen
Within ±7%
Trace Area
Full Screen
Within ±7%
Circ
Full Screen
Within ±10%
Angle
Full Screen
Within ±3%
Volume
Full Screen
Within ±12 %
Table 2 Basic Time Motion Measurements
Parameter
Range
Error
Distance
Full Screen
Within ±4 %
Time
Timeline Display
Within ±2%
Heart Rate
Timeline Display
Within ±5%
Velocity (PW mode)
10-200cm/s (for C5-2s, C11-3s,
L12-4s, L7-3s, L14-6s, L14-
6Ns, V11-3Ws, 7LT4s, P7-3Ts,
L14-5sp, L20-5s, 7L4s, SC6-1s,
6CV1s, L9-3s, C5-1s, C4-1s,
L11-3VNs, L12-3RCs, L14-
5Ws)
10-300cm/s (for P4-2s, P10-4s,
P7-3s, SP5-1s)
When angle≤60º, ≤5%.
Velocity (CW mode)
10-200cm/s (for P7-3Ts, C4-1s)
10-300cm/s (for P4-2s, P10-4s,
P7-3s, SP5-1s)
When angle≤60º, ≤5%. (not including
pencil probe)
NOTE:
Within the selected field range, measurement accuracy is ensured within the range
mentioned above. The accuracy specifications are performed in the worst conditions,
or based on real system tests, regardless of acoustic speed error.
Содержание TE5
Страница 1: ...TE7 TE5 Diagnostic Ultrasound System Operator s Manual Basic Volume ...
Страница 2: ......
Страница 6: ......
Страница 12: ......
Страница 24: ......
Страница 36: ......
Страница 54: ......
Страница 56: ...4 2 Exam Preparation 4 1 1 New Patient Information The Patient Info screen is shown as follows 2 1 3 ...
Страница 110: ......
Страница 115: ...Display Cine Review 6 5 6 Tap Return on the screen or tap Freeze to exit image compare ...
Страница 120: ......
Страница 124: ......
Страница 156: ......
Страница 174: ......
Страница 192: ...12 18 Setup Select Advanced and do as follows Select MAPS and do as follows ...
Страница 202: ...13 2 Probes and Biopsy C5 2s L12 4s L7 3s P4 2s L14 6s C11 3s L14 6Ns V11 3Ws P7 3Ts 7LT4s ...
Страница 203: ...Probes and Biopsy 13 3 7L4s P10 4s L20 5s P7 3s L14 5sp SC6 1s SP5 1s 6CV1s L9 3s C5 1s L11 3VNs C4 1s ...
Страница 222: ...13 22 Probes and Biopsy No Name Description 8 Grooves of the needle guided bracket Matched with the tabs of the probe ...
Страница 226: ...13 26 Probes and Biopsy NGB 034 NGB 035 ...
Страница 250: ......
Страница 272: ......
Страница 276: ...A 4 Wireless LAN Tap Add Manually create a network profile to set ...
Страница 282: ......
Страница 318: ......
Страница 322: ......
Страница 323: ...P N 046 006959 07 1 0 ...