1-4 Safety Precautions
8.
Do not use an aftermarket probe other than those specified
by Mindray. The probes may damage the system, causing a
profound failure, e.g. a fire in the worst case.
9.
Do not subject the probes to knocks or drops. Use of a
defective probe may cause an electric shock.
10.
Do not open the covers and front panel of the system.
Short circuit or electric shock may result when the system
hardware is exposed and powered on.
11.
DO NOT use this system simultaneously with equipment
such as an electrosurgical unit, high-frequency therapy
equipment, or a defibrillator, etc. This would result in a risk
of electric shock to the patient.
12.
When moving the system, you should first fold the LCD
display, disconnect the system from other devices
(including probes) and disconnect the system from the
power supply.
13.
Accessory equipment connected to the analog and digital
interfaces must comply with the relevant IEC standards
(e.g., IEC 60950 information technology equipment safety
standard and IEC 60601-1 medical equipment standard).
Furthermore, all configurations must comply with the
standard IEC 60601-1 chapter 16 ME System. It is the
responsibility of the person, who connects additional
equipment to the signal input or output ports and
configures a medical system, to verify that the system
complies with the requirements of IEC 60601-1 chapter 16
ME System. If you have any questions regarding these
requirements, consult your sales representative.
14.
Prolonged and repeated use of display controls may result
in hand or arm nerve disorders for some individuals.
Observe the local safety or health regulations concerning
the use.
15.
DO NOT contact both the patient and the ultrasound
system or the live parts of the ultrasound system (e.g.
signal I/O ports). Electric shock may occur.
16.
When using intra-cavity transducers, do not activate the
transducer outside the patient’s body.
17.
If you have any doubts of the installation or routing of
external protective cables, use the internal power supply
of the system.
18.
DO NOT put the ultrasound system in any soft materials
(e.g. soft cloth or cotton, etc.) in case that the air vent
becomes blocked.
19.
Always read and follow carefully the manufacturer
instructions on the contrast agent label.
20.
Only use the ECG leads provided with the ECG module;
otherwise electric shock may result.
Содержание TE5
Страница 1: ...TE7 TE5 Diagnostic Ultrasound System Operator s Manual Basic Volume ...
Страница 2: ......
Страница 6: ......
Страница 12: ......
Страница 24: ......
Страница 36: ......
Страница 54: ......
Страница 56: ...4 2 Exam Preparation 4 1 1 New Patient Information The Patient Info screen is shown as follows 2 1 3 ...
Страница 110: ......
Страница 115: ...Display Cine Review 6 5 6 Tap Return on the screen or tap Freeze to exit image compare ...
Страница 120: ......
Страница 124: ......
Страница 156: ......
Страница 174: ......
Страница 192: ...12 18 Setup Select Advanced and do as follows Select MAPS and do as follows ...
Страница 202: ...13 2 Probes and Biopsy C5 2s L12 4s L7 3s P4 2s L14 6s C11 3s L14 6Ns V11 3Ws P7 3Ts 7LT4s ...
Страница 203: ...Probes and Biopsy 13 3 7L4s P10 4s L20 5s P7 3s L14 5sp SC6 1s SP5 1s 6CV1s L9 3s C5 1s L11 3VNs C4 1s ...
Страница 222: ...13 22 Probes and Biopsy No Name Description 8 Grooves of the needle guided bracket Matched with the tabs of the probe ...
Страница 226: ...13 26 Probes and Biopsy NGB 034 NGB 035 ...
Страница 250: ......
Страница 272: ......
Страница 276: ...A 4 Wireless LAN Tap Add Manually create a network profile to set ...
Страница 282: ......
Страница 318: ......
Страница 322: ......
Страница 323: ...P N 046 006959 07 1 0 ...