13-38 Probes and Biopsy
13.2.6 Verifying the Biopsy Guide Line
WARNING:
Prior to each biopsy procedure, be sure to verify the
guide line.
If the needle is not consistent with the guide line, DO
NOT perform the biopsy procedure.
NOTE:
You can perform guide line verification on a single live B/C image, and all biopsy-
irrelevant operations are forbidden.
Adjusting the needle mark is necessary before each biopsy procedure.
1. Confirm that the needle-guided bracket has been installed securely in the correct position.
2. Prepare a container filled with sterile water.
3. Place the head of the probe in the sterile water and place a biopsy needle in the needle guide.
4. When the biopsy needle appears on the image, select [Biopsy] -> [Alignment] to align the
guide line with the biopsy needle.
Tap [Alignment] in the Biopsy menu to enter the Biopsy Verify menu.
Adjust the guide line position
Tap and drag the top dot on the line to change the position of the guide line.
Adjust the angle
Tap and drag the bottom dot on the line to change the guide line angle.
Save the verified settings
After the position and angle of the guide line are adjusted, tap [Save] and the system saves
the current guide line settings. If biopsy is entered again, the displayed Position and Angle are
the verified value.
Restore the factory default settings
Tap [Load Default] and the position and angle of the guide line are restored to the factory
default settings.
Exit biopsy verify status
Tap [Exit] and the system exits the guide line verification status.
Содержание TE5
Страница 1: ...TE7 TE5 Diagnostic Ultrasound System Operator s Manual Basic Volume ...
Страница 2: ......
Страница 6: ......
Страница 12: ......
Страница 24: ......
Страница 36: ......
Страница 54: ......
Страница 56: ...4 2 Exam Preparation 4 1 1 New Patient Information The Patient Info screen is shown as follows 2 1 3 ...
Страница 110: ......
Страница 115: ...Display Cine Review 6 5 6 Tap Return on the screen or tap Freeze to exit image compare ...
Страница 120: ......
Страница 124: ......
Страница 156: ......
Страница 174: ......
Страница 192: ...12 18 Setup Select Advanced and do as follows Select MAPS and do as follows ...
Страница 202: ...13 2 Probes and Biopsy C5 2s L12 4s L7 3s P4 2s L14 6s C11 3s L14 6Ns V11 3Ws P7 3Ts 7LT4s ...
Страница 203: ...Probes and Biopsy 13 3 7L4s P10 4s L20 5s P7 3s L14 5sp SC6 1s SP5 1s 6CV1s L9 3s C5 1s L11 3VNs C4 1s ...
Страница 222: ...13 22 Probes and Biopsy No Name Description 8 Grooves of the needle guided bracket Matched with the tabs of the probe ...
Страница 226: ...13 26 Probes and Biopsy NGB 034 NGB 035 ...
Страница 250: ......
Страница 272: ......
Страница 276: ...A 4 Wireless LAN Tap Add Manually create a network profile to set ...
Страница 282: ......
Страница 318: ......
Страница 322: ......
Страница 323: ...P N 046 006959 07 1 0 ...