5-42 Image Optimization
Description of parameters:
Parameter
Description
Method
Function: select the image acquisition method.
Selection: Rocked, Linear.
Linear mode: during the sweep, the probe must be kept parallel. The
scanning speed should be constant.
Rocked mode: in this mode, the probe must be moved to a position
where you can clearly see a middle cut of the object you want to scan
and render. Tilt the probe to about 30 degrees until the object you
want to scan disappears. Start the acquisition and tilt the probe over a
distance of around 60 degrees until the object disappears again.
During the sweep, the probe may not be moved parallel, just tilted.
Tip: the speed is related to scanning distance or angle.
Distance
Function: to set the distance the probe covered from one end to the
other end during a linear sweep.
Range: 10-200 mm, in increments of 10 mm.
Angle
Function: to set the angle the probe covered during a fan sweep.
Range: 10-80°, in increments of 2°.
Render
Mode
Surface
Function: set Surface as the 3D image rendering mode.
This is useful for surface imaging, such as fetus face, hand or foot.
Tip: you may have to adjust the threshold to obtain a clear body
boundary.
Max.
Function: set Max. as the 3D image rendering mode. Displays the
maximum echo intensity in the observation direction.
This is useful for viewing bony structures.
Min.
Function: set Min. as the 3D image rendering mode. Displays the
minimum echo intensity in the observation direction.
This is useful for viewing vessels and hollow structures.
X-ray
Function: set X-ray as the 3D image rendering mode. Displays the
average value of all gray values in the ROI.
X Ray: used for imaging tissues with different internal structures or
tissues with tumors.
5.11.2.3 Smart 3D Image Viewing
Enter/Exit Image Viewing
To enter image viewing:
The system enters image viewing when image acquisition is complete.
Exit
To return to Smart 3D image acquisition preparation status, tap [Update] or [Freeze].
Activate MPR
Tap [A], [B], [C] or [VR] to activate MPR or 3D image (VR).
Содержание TE5
Страница 1: ...TE7 TE5 Diagnostic Ultrasound System Operator s Manual Basic Volume ...
Страница 2: ......
Страница 6: ......
Страница 12: ......
Страница 24: ......
Страница 36: ......
Страница 54: ......
Страница 56: ...4 2 Exam Preparation 4 1 1 New Patient Information The Patient Info screen is shown as follows 2 1 3 ...
Страница 110: ......
Страница 115: ...Display Cine Review 6 5 6 Tap Return on the screen or tap Freeze to exit image compare ...
Страница 120: ......
Страница 124: ......
Страница 156: ......
Страница 174: ......
Страница 192: ...12 18 Setup Select Advanced and do as follows Select MAPS and do as follows ...
Страница 202: ...13 2 Probes and Biopsy C5 2s L12 4s L7 3s P4 2s L14 6s C11 3s L14 6Ns V11 3Ws P7 3Ts 7LT4s ...
Страница 203: ...Probes and Biopsy 13 3 7L4s P10 4s L20 5s P7 3s L14 5sp SC6 1s SP5 1s 6CV1s L9 3s C5 1s L11 3VNs C4 1s ...
Страница 222: ...13 22 Probes and Biopsy No Name Description 8 Grooves of the needle guided bracket Matched with the tabs of the probe ...
Страница 226: ...13 26 Probes and Biopsy NGB 034 NGB 035 ...
Страница 250: ......
Страница 272: ......
Страница 276: ...A 4 Wireless LAN Tap Add Manually create a network profile to set ...
Страница 282: ......
Страница 318: ......
Страница 322: ......
Страница 323: ...P N 046 006959 07 1 0 ...