GE
DRAFT
LOGIQ P9/P7
D
IRECTION
5604324
, R
EVISION
11
DRAFT (J
ANUARY
24, 2019)
S
ERVICE
M
ANUAL
Chapter 3 - Setup Instructions
3-15
3-6-3
Connecting the Secondary “Patient” LCD Monitor
1) Power OFF/Shutdown the system as described in:
2) Connect the Secondary Monitor according to connection scheme (see:
).
CAUTION
!! CAUTION:
Secondary LCD “Patient” Monitor MUST NEVER be connected to the LOGIQ P9/P7 ultrasound
systems mains supply directly!
Always connect it to the supplied Isolation Transformer!
The Secondary Monitor is the only item to be connected to the Transformer.
NOTICE
!! NOTICE:
Secondary Monitor is
NOT intended for diagnostic use
.
It is an additional device used to allow the patient to watch the proceedings.
Take your time to think about the best position of the monitor in your facilities.
Patients should be able to view the monitor easily and without having to bend or turn around.
NOTICE
!! NOTICE:
Connection to Secondary Monitor is via HDMI cable, but HDMI connector output does NOT carry audio
signal.