GE
RAFT
LOGIQ P9/P7
D
IRECTION
5604324
, R
EVISION
11
DRAFT (J
ANUARY
24, 2019)
S
ERVICE
M
ANUAL
8-108
Section 8-17 - Replacement around Nest Box
8-17-9
Removal Procedure - Separating DCWD Board(Option) from MPI
Philips screw-driver required to add this part.
1) Remove 3 screws on MPI assy.
Figure 8-118 DCWD Board screw points
2) Carefully separate DCWD board from MPI assy not to damage the DCWD connector. Refer to
following figure
.
Figure 8-119 Position of DCWD board connector