Section 5-11 - Common Service Desktop (R3 only)
To remotely start, restart, or shut down the
LOGIQ P9/P7:
1.) Navigate to
Utility (second page) > Service > Utilities > Disruptive Mode Utility
and set
Disruptive mode to On.
2.) Navigate to
Utility (second page) > Service > Utilities > System Shutdown
and then select an
operation:
-
To shut down the LOGIQ P9/P7, select
Shutdown System
.
-
To restart the LOGIQ P9/P7, select
Restart System
.
-
To shut down the application software, select
Shutdown Application
.
3.) To retain the current setting for Disruptive mode, select
Retain Disruptive Mode
.
4.) Click
Submit Request
.
For more information, see
:
•