Chapter 7 - Diagnostics/Troubleshooting
7-11
7-4-3-4
Hardware configuration error
7-4-3-4-1
Cause
Hardware(s) not detected as valid system. System will boot up but will not allow scanning.
7-4-3-4-2
What to do
1) Press OK, and shutdown the system and reboot.
2) If same issue persists, then shut down the system, system breaker.
7-4-3-4-3
Log message of the Invalid hardware error
Example
Error ; acqHWCAP(1776); ********************************************
Error ; acqHWCAP(1776); *** ***
Error ; acqHWCap(1776); *** HW RACK CONFIGURATION INVALID!!! ***
Error ; acqHWCAP(1776); *** ***
Error ; acqHWCAP(1776); ********************************************
Error ; acqHWCAP(1776); *** ***
Error ; acqHWCAP(1776); *** Board FuncRevs Provided ***
Error ; acqHWCAP(1776); *** ----------------------- ***
Error ; acqHWCAP(1776); *** MCB_FuncRev1A (qty=1)
Error ; acqHWCAP(1776); *** COMEXP_FuncRev1A (qty=1)
Error ; acqHWCAP(1776); *** MDC_FuncRev1A (qty=1)
Error ; acqHWCAP(1776); *** MPI_FuncRev1A (qty=1)
Error ; acqHWCAP(1776); *** MIO_FuncRev1A (qty=1)
Error ; acqHWCAP(1776); *** ***
Error ; acqHWCAP(1776); ********************************************
Info ; DFEManager(1776); HandleRunFlag: setting RunDfe and RunBlock flags to true!
Error ; EAErrorHandler(1776); Error from acqHWCAP: Invalid Hardware Rack Configuration - Check
logs. Severity : FatalHW
Содержание LOGIQ P9
Страница 2: ......
Страница 11: ...GE LOGIQ P9 P7 DIRECTION 5604324 REVISION 11 SERVICE MANUAL ix ZH CN KO ...
Страница 40: ...GE LOGIQ P9 P7 DIRECTION 5604324 REVISION 11 SERVICE MANUAL xxxviii Table of Contents ...
Страница 184: ...GEDRAFT LOGIQ P9 P7 DIRECTION 5604324 REVISION 11 DRAFT JANUARY 24 2019 SERVICE MANUAL Chapter 4 Functional Checks 4 44 ...
Страница 601: ...This page was intentionally left blank ...