Disable Remote Connectivity
This utility will uninstall remote connectivity items like Agent Configuration, SSH and Virtual Console
Observation (VCO) from Service Desktop. If remote connectivity is disabled, then the Submit Changes
button on Agent Configuration and the Start/Stop buttons for SSH and VCO will all be disabled. In
addition, some scripts will execute to make the LOGIQ E10 more secure.
NOTE:
If you have disabled remote connectivity and then performed a software reload, you will have
to disable remote connectivity again.
To access Disable Remote Connectivity, navigate to Utility (second page) > Service > Utilities, and then
select Disable Remote Connectivity.
Figure 5-52 Disable Remote Connectivity
To disable remote connectivity:
1.) Navigate to Utility (second page) > Service > Utilities, and then select Disable Remote Connectivity.
2.) Select Disable Remote Connectivity.
3.) To disable connectivity to the LOGIQ E10 through Agent Configuration, SSH, and Virtual Console
Observation, click OK.
4.) Once connectivity has been disabled, select OK again.
Содержание LOGIQ P9
Страница 2: ......
Страница 11: ...GE LOGIQ P9 P7 DIRECTION 5604324 REVISION 11 SERVICE MANUAL ix ZH CN KO ...
Страница 40: ...GE LOGIQ P9 P7 DIRECTION 5604324 REVISION 11 SERVICE MANUAL xxxviii Table of Contents ...
Страница 184: ...GEDRAFT LOGIQ P9 P7 DIRECTION 5604324 REVISION 11 DRAFT JANUARY 24 2019 SERVICE MANUAL Chapter 4 Functional Checks 4 44 ...
Страница 601: ...This page was intentionally left blank ...