GE
RAFT
LOGIQ P9/P7
D
IRECTION
5604324
, R
EVISION
11
DRAFT (J
ANUARY
24, 2019)
S
ERVICE
M
ANUAL
8-152
Section 8-29 - Replacement of Peripheral Cable Assy
8-29-2-5
Installation Procedure
Parts to be installed in reverse order of removal. But, refer to below procedure for cable ties.
Figure 8-179 Tie cables with the tie through the holes
8-29-2-6
Functional Check
Test 1: Print an image by pressing the button P1 on the operator panel. (P1 key should be configured
correctly. Refer to
Adding Printer to the system)
Verify1: - Verify that image can be printed.
Verify2: - Verify that printed image has no distortion as compared the image on screen.
2 holes for fixing cable tie
Cable tie
Align the head of cable tie as shown