Chapter 8 - Replacement Procedures
8-131
NOTE:
There are various types of VPD EEPROM integrated in the system.
8-23-6-1
Accessing EEPROM
1) Exit to Desktop (Refer to
for details)
2) Start VPD Editor by clicking icon on the desktop. (“Vpdedit”)
3) Select COMEXP from device grid.
Board Name
Main VPD Contents
Replaceable
Physical VPD location
MCB
FuncRev
No
Integrated with MCB
MPI
FuncRev
No
Integrated with MPI
MDC
FuncRev
No
Integrated with MDC
COMEXP
FuncRev, , SSN (System Serial Number
Yes
On MCB board as replaceable
EEPROM.
MIO
FuncRev
No
Integrated with MIO
DC4D
FuncRev
No
Integrated with DC4D
OPIOR
FuncRev
No
Integrated with OPIO