Section 5-11 - Common Service Desktop (R3 only)
To perform a network capture
:
1.) Navigate to
Utility > Service > Utilities > Network Capture
.
2.) From
Network Capture
, do the following:
-
Under
Maximum Size
, select the allowed size of the generated log file.
-
Under
Devices
, select the DICOM-configured device for which you want to capture
information. If no additional devices are configured, only
All
will be available.
3.) Select
Start Network Capture
to start the process. This causes the network capture to start,
enables the
Stop
button, and updates the
Network Capture Status
pane and changes the
Status
to
Running
.
4.) Click the
Stop
button to end data collection. Stopping is a two-step process:
-
Stops the data collection and immediately closes the .etl file.
-
Collects additional diagnostic data that may help diagnose network issues. When the file is
closed, you see “There is no trace session currently in progress”. When the remaining data is
collected and the .cab file is closed, you are notified in the banner.
For more information, see
:
•
Содержание LOGIQ P9
Страница 2: ......
Страница 11: ...GE LOGIQ P9 P7 DIRECTION 5604324 REVISION 11 SERVICE MANUAL ix ZH CN KO ...
Страница 40: ...GE LOGIQ P9 P7 DIRECTION 5604324 REVISION 11 SERVICE MANUAL xxxviii Table of Contents ...
Страница 184: ...GEDRAFT LOGIQ P9 P7 DIRECTION 5604324 REVISION 11 DRAFT JANUARY 24 2019 SERVICE MANUAL Chapter 4 Functional Checks 4 44 ...
Страница 601: ...This page was intentionally left blank ...