Chapter 5 - Components and Functions (Theory)
5-55
This table shows all the elements available on
Bootup Issues
with descriptions.
For more information, see
:
•
•
Table 5-20 Bootup Issues
Element
DESCRIPTION
ID
Identifier for the system software boot. For example, 10 for the tenth boot and so on.
Timestamp
Date and time that the associated issue for the boot occurred.
Cycle Count
Number of times EchoLoader was restarted during the associated boot.
• 0 - Normal. EchoLoader started without attempting a restart.
• 1 or more - Indicates the number if times EchoLoader attempted to restart.
Missing Devices
Displays the devices that were missing during the boot.
Incompatible Board
Displays whether an incompatible board exists.
Battery State ID
Type of battery detected. For example,
EPS
or
PA
.
Invalid
displays if no battery was
detected.
Previous Boot Crash Dump
Displays whether a crash dump has occurred. Valid values are:
• 1 - Indicates a crash dump has occurred in the previous boot.
• 0 - Normal. No crash dump has occurred.
Warm Boot
Displays whether a warm boot has occurred. Valid values are:
• 1 - The system was rebooted.
• 0 - The system was started after a shutdown.
AC Fail
Displays whether an AC failure occurred during the boot. Valid values are:
• 1 - There was an AC failure in the boot.
• 0 - Normal. No AC failure has occurred.
Abnormal Context
Displays whether the software has booted to a stable state (Interface manager
successfully initiated). Valid values are:
• 1 - The system has not booted to a stable state. Only abnormal logs display.
• 0 - Normal. The system initialized to a stable state.
Exit Requested
Displays the type of exit that has been requested. For example, 4 displays the
operating system was restarted.
Echoloader Exit Code
Displays the exit code generated by EchoLoader. 0 indicates that Echoloader exited
without any issues.
Echoloader Exit String
Displays the type of exit requested by EchoLoader. For example, restart the
operating system or exit the application software.
Содержание LOGIQ P9
Страница 2: ......
Страница 11: ...GE LOGIQ P9 P7 DIRECTION 5604324 REVISION 11 SERVICE MANUAL ix ZH CN KO ...
Страница 40: ...GE LOGIQ P9 P7 DIRECTION 5604324 REVISION 11 SERVICE MANUAL xxxviii Table of Contents ...
Страница 184: ...GEDRAFT LOGIQ P9 P7 DIRECTION 5604324 REVISION 11 DRAFT JANUARY 24 2019 SERVICE MANUAL Chapter 4 Functional Checks 4 44 ...
Страница 601: ...This page was intentionally left blank ...