Chapter 8 - Replacement Procedures
8-29
11) A confirmation window appears, and then hit the Enter key.
12) Approx. 40 minutes later, loading the software will be completed.
13) Remove Base System Software DVD, and type "exit" to reboot the system.
14) At the first system reboot after loading the software, the OS will reboot the system automatically.
15) After the reboot, the installation of system (Windows Embedded Standard
will be fully completed.
16) Go to BIOS mode (Refer to the procedure 2) 3) above.)