GE
RAFT
LOGIQ P9/P7
D
IRECTION
5604324
, R
EVISION
11
DRAFT (J
ANUARY
24, 2019)
S
ERVICE
M
ANUAL
3-26
Section 3-7 - Completing the Set Up
•
Logoff
Use this button to log off the current user. The system remains ON and ready for a new user to log
on. If the Logoff button is dimmed, it indicates that no user is logged on to the unit at the moment.
•
Shutdown
Use this button to shut down the system. The entire system will shut down. It is recommended to
perform a full shutdown at least once a week.
If the Shutdown button is dimmed, press the power button or Alt F10 to shut down the unit.
•
Cancel
Use this button to exit from the System-Exit menu and return to the previous operation.
•
Exit
(Only available when logged in as GE Service with Service Dongle)
Select this button when you want to exit to the Windows Desktop.
•
Sleep (R2 and R2.5 Only)
Use Sleep Mode when you do a portable exam in order to reduce the time to start up the system.
When you use Sleep Mode, it takes ~90 seconds to start up the system versus 2-3 minutes.
•
Change Password
Use this button to change User Password.
Section 4-2-5-4 "Sleep Mode Check" on page 4-14
NOTICE
!! NOTICE:
After turning off a system, wait at least 10 seconds before turning it on again.
The system may not be able to boot if power is recycled too quickly.
Содержание LOGIQ P9
Страница 2: ......
Страница 11: ...GE LOGIQ P9 P7 DIRECTION 5604324 REVISION 11 SERVICE MANUAL ix ZH CN KO ...
Страница 40: ...GE LOGIQ P9 P7 DIRECTION 5604324 REVISION 11 SERVICE MANUAL xxxviii Table of Contents ...
Страница 184: ...GEDRAFT LOGIQ P9 P7 DIRECTION 5604324 REVISION 11 DRAFT JANUARY 24 2019 SERVICE MANUAL Chapter 4 Functional Checks 4 44 ...
Страница 601: ...This page was intentionally left blank ...