10-28
Section 10-8 - Electrical Safety Tests
10-8-6-2
Generic Procedure
The testing meter is connected from accessible protectively-earthed parts of the equipment to ground.
Measurements should be made with power line polarity normal and reversed, and with the neutral open
and closed. Record the highest reading of current.
NOTE:
Color of Power outlet cable
i) USA
and
Canada
Hot: Black
Neutral: White
Ground: Green or Green-yellow
ii) Others
Hot: Brown
Neutral: Blue
Ground: Green-yellow
When using Dale 600/601 or a similar test instrument, its power plug may be inserted into the wall outlet
and the equipment under test is plugged into the receptacle on the panel of the meter. This places the
meter in the grounding conductor and the current flowing through protective earth will be indicated in
any of the current ranges. The maximum allowable limit for earth leakage current is shown in
Figure 10-19 Set Up for Test of Earth Leakage Current, UL60601-1/IEC 60601-1 Clause 19
Table 10-17 Typical Data Sheet for Earth Leakage Current
Tester Neutral
Tester Polarity Switch
Measured Earth leakage current
Enter Name of tested peripheral here:
Closed
Normal
Closed
Reversed
Open
Normal
Open
Reversed
Содержание LOGIQ P9
Страница 2: ......
Страница 11: ...GE LOGIQ P9 P7 DIRECTION 5604324 REVISION 11 SERVICE MANUAL ix ZH CN KO ...
Страница 40: ...GE LOGIQ P9 P7 DIRECTION 5604324 REVISION 11 SERVICE MANUAL xxxviii Table of Contents ...
Страница 184: ...GEDRAFT LOGIQ P9 P7 DIRECTION 5604324 REVISION 11 DRAFT JANUARY 24 2019 SERVICE MANUAL Chapter 4 Functional Checks 4 44 ...
Страница 601: ...This page was intentionally left blank ...