5-56
Section 5-11 - Common Service Desktop (R3 only)
Alert Logs
Alert Logs
displays the details for each alert. Alerts are status logs typically reported as a result of an
error. Alerts are sent to the back office to be managed and reported to service.
To access this page, under
System Status
, select
Alert Logs
.
Figure 5-38 Alert Logs
This table shows all the elements available on
Alert Logs
with descriptions.
For more information, see
:
•
•
Table 5-21 Alert Logs
Element
DESCRIPTION
Alert ID
Identifier for the alert log that has been sent to the back office.
Sub System
Hardware component.
Rule
Code for the alert. Alerts are filtered in the back office using rule codes.
Description
Description of the alert log.
Value
Current value read for the associated hardware component.
Lower Limit
Value for the lower limit for the associated hardware component.
Upper Limit
Value for the upper limit for the associated hardware component.
Severity
Severity of the alert. These are the valid values:
• 1 - Error. Out of specification
• 2 - Warning
• 3 - Normal
First Alert Time
Time the alert first occurred.
Last Alert TIme
Time the alert last occurred.
Last Alert Local Time
Time (local) the alert last occurred.
Last Alert Sent Time
Time the alert was last sent to the back office.
Count
Number of times that the alert has occurred.
Содержание LOGIQ P9
Страница 2: ......
Страница 11: ...GE LOGIQ P9 P7 DIRECTION 5604324 REVISION 11 SERVICE MANUAL ix ZH CN KO ...
Страница 40: ...GE LOGIQ P9 P7 DIRECTION 5604324 REVISION 11 SERVICE MANUAL xxxviii Table of Contents ...
Страница 184: ...GEDRAFT LOGIQ P9 P7 DIRECTION 5604324 REVISION 11 DRAFT JANUARY 24 2019 SERVICE MANUAL Chapter 4 Functional Checks 4 44 ...
Страница 601: ...This page was intentionally left blank ...