Section 8-13 - Replacement of Fixed Arm Cover and Fixed Arm Holder Grip
Section 8-13
Replacement of Fixed Arm Cover and Fixed Arm Holder Grip
8-13-1
Manpower
2 person, 5 minutes.
8-13-2
Tools
Standard Phillips Screwdriver, Hex Wrench.
8-13-3
Removal Procedure
8-13-3-1
Removal Procedure - Fixed Arm Cover
1) Remove 1 screw and lift up the Fixed Arm Cover.
Figure 8-70 Fixed Arm Cover screw point
8-13-3-2
Removal Procedure - Fixed Arm Holder Grip
1) Remove 2 screws and pull out the Fixed Arm Holder Grip.
Figure 8-71 Fixed Arm Holder Grip Cover screw points