GE
DRAFT
LOGIQ P9/P7
D
IRECTION
5604324
, R
EVISION
11
DRAFT (J
ANUARY
24, 2019)
S
ERVICE
M
ANUAL
Chapter 8 - Replacement Procedures
8-9
8-3-5-2
Preparation for Software Loading
8-3-5-2-1
Recording important settings and parameters (optional)
If the system has not been deployed into a customer site yet (i.e. the system has been in the factory
default and there has not been patient data), this step can be skipped.
8-3-5-2-1-1
Overview
See the following pages to record data.
These subsections include descriptions for recording/setting up data for the following screens.
With the system Powered On, follow the procedures below to document basic system information and
back up Presets, and if required, Patient Archive and Patient Images prior to software load.
•
Utility -> Connectivity
-
TCP/IP (see: “Connectivity - Recording the TCP/IP settings”).
-
Device (see: “Connectivity - Recording the AE Title and Port settings”).
-
Miscellaneous (see: “Connectivity - Recording Miscellaneous settings”).
•
Utility -> Admin
-
System Admin (see: “Admin - Recording the Software Option Keys”).
These subsections include descriptions for recording/setting up data for the following screens.
Hardcopy forms are provided here for convenience, but content will be included if selected during
Backup process. Backing up this information is only required if an upgrade specifies the information be
backed up or as an extra precaution against loss due to a corrupt back-up disk.
•
Utility -> Connectivity
-
Device (see: “Connectivity - Recording Device Destination settings”).
-
Service (see: “Connectivity - Recording Service settings”).
-
Dataflow (see: “Connectivity - Recording Dataflow settings”).
-
Buttons (see: “Connectivity - Recording the Print Key Assignments”).
-
Removable Media (see: “Connectivity - Service settings”).
•
Utility -> Admin
-
Users (see: “Admin - Users”).
•
Utility -> System
-
Backup/Restore (see: “System - Data Store Management”).
-
Peripherals (see: “System - Recording Peripheral settings”).
-
About (see: “System - Recording software and hardware versions”).
-
General, Record preset Region, Language and Regional Options. (See the latest revision
ofthe LOGIQ P9/P7 Basic User Manual))
8-3-5-2-1-2
Connectivity - Recording the TCP/IP settings
1) From the touch panel, select Utility -> Connectivity -> TCP/IP.
Содержание LOGIQ P9
Страница 2: ......
Страница 11: ...GE LOGIQ P9 P7 DIRECTION 5604324 REVISION 11 SERVICE MANUAL ix ZH CN KO ...
Страница 40: ...GE LOGIQ P9 P7 DIRECTION 5604324 REVISION 11 SERVICE MANUAL xxxviii Table of Contents ...
Страница 184: ...GEDRAFT LOGIQ P9 P7 DIRECTION 5604324 REVISION 11 DRAFT JANUARY 24 2019 SERVICE MANUAL Chapter 4 Functional Checks 4 44 ...
Страница 601: ...This page was intentionally left blank ...