Chapter 5 - Components and Functions (Theory)
5-83
Reset Patient Database
Reset Patient Database
resets the LOGIQ P9/P7 patient database. This operation automatically shuts
down the application before cleaning the patient database. Once the operation is complete, the LOGIQ
P9/P7 reboots with an empty database. Disruptive mode must be On to reset the patient database. If
Disruptive mode is Off, a warning message displays asking to turn on Disruptive mode.
Reset Patient Database
deletes the contents of the following folders:
•
E:\Scanner\GEMS_IMG
•
E:\Scanner\GEMS_REP
The information on
Reset Patient Database
is available to Class C (Service Advanced) and Class M
licenses.
To access
Reset Patient Database
, select
Utility (second page) > Service > Utilities > Reset
Patient Database
.
Figure 5-56 Reset Patient Database
This table shows all the elements available on
Reset Patient Database
with descriptions.
To reset the patient database
:
1.) Back up the patient database.
2.) Navigate to
Utility (second page) > Service > Utilities > Disruptive Mode Utility
and set
Disruptive mode to On.
3.) Navigate to
Utility (second page) > Service > Utilities > Reset Patient Database
.
4.) Click
Reset Patient Database
.
5.) Confirm the patient database has been emptied.
For more information, see
:
•
Table 5-35 Reset Patient Database
Element
DESCRIPTION
Instruction
Information about the database reset.
Reset Patient Database
When Disruptive mode is ON, resets the LOGIQ P9/P7 patient database.
Содержание LOGIQ P9
Страница 2: ......
Страница 11: ...GE LOGIQ P9 P7 DIRECTION 5604324 REVISION 11 SERVICE MANUAL ix ZH CN KO ...
Страница 40: ...GE LOGIQ P9 P7 DIRECTION 5604324 REVISION 11 SERVICE MANUAL xxxviii Table of Contents ...
Страница 184: ...GEDRAFT LOGIQ P9 P7 DIRECTION 5604324 REVISION 11 DRAFT JANUARY 24 2019 SERVICE MANUAL Chapter 4 Functional Checks 4 44 ...
Страница 601: ...This page was intentionally left blank ...