Section 7-7 - Troubleshooting Trees, Instructions and Tech Tips
To view logs in the ScLogWindow:
1.) With the ScLogWindow open, double-click the
EchoLoader
icon.
2.) Use
Alt+Tab
to put the ScLogWindow in front.
3.) While the application is booting up, observe logging in ScLogWindow.
To save logs in ScLogWindow:
•
Select
File
>
Save as
To filter logs in ScLogWindow:
1.) Select
View
and check the entries to filter.
2.) Select
Stop
to stop the logging.
3.) View the filtered logs.
4.) To clear the window and start a fresh log capture, select
Clear
.