GE
DRAFT
LOGIQ P9/P7
D
IRECTION
5604324
, R
EVISION
11
DRAFT (J
ANUARY
24, 2019)
S
ERVICE
M
ANUAL
Chapter 5 - Components and Functions (Theory)
5-85
To clean user-defined data
:
1.) If you want to use the current settings later, back them up now.
2.) Navigate to
Utility (second page) > Service > Utilities > Disruptive Mode Utility
and set
Disruptive mode to On.
3.) Navigate to
Utility (second page) > Service > Utilities > Clean Userdefs
.
4.) Select the data you want to set to the default settings.
5.) Click
Clean Userdefs
.
For more information, see
:
•
Содержание LOGIQ P9
Страница 2: ......
Страница 11: ...GE LOGIQ P9 P7 DIRECTION 5604324 REVISION 11 SERVICE MANUAL ix ZH CN KO ...
Страница 40: ...GE LOGIQ P9 P7 DIRECTION 5604324 REVISION 11 SERVICE MANUAL xxxviii Table of Contents ...
Страница 184: ...GEDRAFT LOGIQ P9 P7 DIRECTION 5604324 REVISION 11 DRAFT JANUARY 24 2019 SERVICE MANUAL Chapter 4 Functional Checks 4 44 ...
Страница 601: ...This page was intentionally left blank ...