Chapter 4 - Functional Checks
4-31
4) Verify BF Mode.
Figure 4-31 B-Flow Mode
5) From the touch panel, select “COLOR”
6) Verify B-Flow Color is shown.
Figure 4-32 B-Flow Color Mode
4-3-7-8
CW Mode (Optional)
NOTE:
Make sure to select probe which supports CWD
1) Press B Mode button.
2) Press CW Mode button.
3) Verify that system CW scan with the activated probe.